Medical Vacuum Suction System Medical Equipment Factory
1. ETR Medical Vacuum Suction Station Description
Medical vacuum is a dedicated to :
- Ensure the surgical drainage of pleural or pericardial effusions for example
- Declutter the airways or gastric patients by aspiration
ETR Medical Vacuum Systems are designed for the central supply of medical vacuum in hospitals/clinics, and in particular in ICU, emergency departments as well as operating theaters. They are designed in accordance with international standards.
Technology & design
ETR has been supplying Medical Vacuum Systems to hospitals and clinics around the world for over 20 years. Featuring high efficiency lubricated or oil free rotary vanes pumps, they come in various available configurations to suit the needs of all healthcare facilities :
- Compact horizontal: integrated on horizontal receiver
- Compact vertical: integrated on a vertical steel frame
- Modular: separated modules to fit specific requirements
- In skids or shelters, according to customers' specifications
ETR medical vacuum systems are composed of :
- a single, double, triple and even quadruple vacuum pump in accordance with the configuration required and local standards
- a medical grade tank with equipment
- a CPU control box
- interconnexions
- an antibacterial filtration
2.ETR Medical Vacuum Suction Station Advantage
- Dynamic control of water cooling temperature and water level, ensuring the operation of vacuum pump in the optimal condition;
- Automatic switching between units, and automatic supplement with the subsequent unit in accordance with the decline in gas consumption;
- UV disinfection and sterilization, achieving sterile and environmental emissions;
- Automatic electrical control system with high degree of intelligence;
- Dynamic display of technological process, enabling operating data storage, fault protection and abnormity warning;
- Open data communication interface, facilitating remote management of equipment.
3. ETR Medical Vacuum Suction Station Working Principle
4. Medical Vacuum Suction Station Specificatons
Item |
Specification |
Manufacturer |
ETR |
Model |
ETV-07 |
Flow Capacity |
4.66Nm3/min |
Vacuum pressure |
-0.02~-0.07Mpa |
Motor Power(KW) |
7.5Kw |
Voltage |
230V,60Hz,3Ph |
Size |
1,300*800*1,500(mm) |
Weight |
900Kg |
Noise Level |
69dB(A) |
5. Medical Vacuum Suction System Model For Selection
Item |
Model |
Capacity
(Nm3/h) |
Pressure
mbar(Mpa) |
Size (mm) |
Weight |
Voltage
(V) |
Power
(kw) |
Noise dB(A) |
L |
W |
H |
(KG) |
1 |
ETV-02 |
80 |
33mbar
(-0.098Mpa) |
1300 |
800 |
1500 |
850kg |
380V |
2.35 |
66 |
2 |
ETV-03 |
110 |
1300 |
800 |
1500 |
900kg |
380V |
3.85 |
72 |
3 |
ETV-04 |
165 |
1250 |
950 |
1800 |
1180kg |
380V |
4 |
63 |
4 |
ETV-05 |
230 |
1250 |
950 |
1800 |
1240kg |
380V |
5.5 |
68 |
5 |
ETV-07 |
280 |
1500 |
1260 |
1900 |
1280kg |
380V |
7.5 |
69 |
6 |
ETV-11 |
400 |
1500 |
1260 |
1900 |
1390kg |
380V |
11 |
73 |
7 |
ETV-15 |
500 |
2100 |
1300 |
1950 |
1800kg |
380V |
15 |
74 |
6. Warranty &Service
Warranty
The Guarantee/Warranty Period shall be a period of twelve months after on-site startup & commissioning or eighteen months after shipment, whichever occurs first. If any trouble or defect, originating with the design, material, workmanship or operating characteristics of any Goods, arises at any time during warranty period, ETR shall, at his own expense and as promptly as possible, make such alterations, repairs and replacements.
On-Site Support
ETR can do paid services of on-site startup, commissioning, installation supervision, training, by providing purchaser with the services of qualified English-speaking
engineer at site. ETR shall obtain all permits and licenses required to perform the services under this Agreement.
COMPANY OVERVIEW
1.Basic Information about ETR
Engineering site