Customization: | Available |
---|---|
Color: | White |
Application: | Medical |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
A medical gas zone valve box is a device that is used to control the flow of medical gases to different areas of a healthcare facility. The box is typically installed near the main medical gas supply source and contains several individual valves that can be used to control the flow of medical gases to different parts of the facility. This allows healthcare providers to manage the distribution of medical gases and ensure that they are delivered to the right areas at the right time. In addition to controlling the flow of medical gases, medical gas zone valve boxes may also be used to monitor the pressure and flow rate of the gases, as well as to provide alarms in the event of a malfunction or other emergency.
Medical gas four safety principles:
Continuous gas supply
Sufficient flow and pressure
Gas identification
Gas Quality
Zone valve box shall be Eter series.
* Each recessed/exposed zone valve shall consist of a steel valve box
constructed of 1.5 mm steel sheet, complete with a backed enamel
finish RAL 9016, which can house single or multiple shut-off ball
valves with tube extensions.
* Valves shall be a 3-piece forged brass and a 4-bolt design, bronze or
brass body, stem, ball bolts, and nuts shall be made of stainless steel,
double seal, full port union ball type, with (RPTFE) seats and Vi-ton
seals, with a pressure rating of 4080 kPa (600 psi) or 40.8 bars.
* Valves shall be operated by a lever-type handle requiring anly a quarter
turn from a fully opened to a fully closed position. All valves shall be
equipped with type "K" or "L" washed and degreased copper pipe stub
extensions of sufficient length to protrude beyond the sides of the box.
* All pipe stub extensions shall be supplied with suitable plugs or caps to
prevent contamination of the assembly prior to installation.
* Each valve shall be supplied with a medical gas identification label as request.
* Valves shall be available with or without line pressure gauges, as required.
Gauges, shall be 50 mm diameter.
* Pressure gauges shall read 0-1100 kPa (0-11 bar) for all gauges except
Nitrogen, which shall read 0-2500 kPa (0-25 bar), and Vacuum, which
shall read 0-100 kPa (0-760 mmHg).
* There are two front options: an aluminum door or a pull-out removable
window.
Medical Gas Supply System Pipelines
Pipe Material |
The main oxygen pipeline and the branch pipeline into the room are used of defatted purple copper tube or stainless steel tube. The pipeline inside of bed head unit is using degreased copper pipe.
|
|
Pipe Setting |
The gas riser pipe is designed in the pipeline well. Pipes should be piped when crossing the wall. The pipes installation should be neat and beautiful.
|
|
Pipe Connection |
Pipe system using welded connection. Metal seals throughout the pipe system connection Horizontal pipe and branch pipe of bed head unit should be with maintenance valve
|
|
Pipe Grounding |
Medical gas pipelines in medical rooms should be equipotential grounded; mainfold/pipelines etc should be anti-static grounded; Independent grounding resistance should less than 10Ω;
|
|
Pipeline Leakage Rate |
Leakage test of medical gas pipelines when no terminal components are connected, the hourly leakage rate should not exceed 0.05%. According to Chinese standard GB50751-2012 |
Zone Valve Box |
Gas Regulator Box |
Monitoring alarm box
|
1,Used to control the on and off of medical gases in the area; 2,1~6 channels of medical gas can be controlled according to user requirements; 3,Silver and copper welding to prevent leakage; 4,Transparent window for easy operation and maintenance; 5,Each gas is marked with a national standard color for easy identification. |
1,Used to provide a pressure-stable source of gas for the clinic; 2,With dual circuit design, it can be maintained without stopping the gas; 3,Use a large flow reducer to ensure pressure stability; 4,Emergency door lock device to close the valve in an emergency; 5,Transparent window for easy operation and maintenance; |
1,Monitor the pressure of the gas supply, and the pressure is high or low to alarm; 2,Monitor the moisture content/oxygen concentration/CO content in oxygen , giving alarm against any deviation from standard set value. Powerful data post-processing function, real-time storage history; 3,intelligent monitoring, with remote data transfer function; |
Medical Gas Supply System Gas Management
The pressure monitoring alarm box is used to monitor the pipeline pressure of the medical center air supply system and respond to pressure changes through a digital microprocessor. When the pressure is higher or lower than the set value, the system will automatically alarm by sound and light, etc., to remind the management to timely process the medical center air supply system.
Comprehensive data collection: monitoring of the source gases and gas-consuming areas in the hospital.
Central monitoring: central monitoring of gas sources and gas-consuming areas, enabling comprehensive data browsing and operation curve analysis.
Fault early warning: abnormal data analysis and trend analysis.
Q: Are you factory or trading company?
A: We are factory specilized in making medical equipments from year 2003.
Q: Where are you?
A: We are in ChangSha,Hunan province,3 hours from guangzhou by high-speed rail.
Q: Do you accept the small order?
A: Yes, you can contact with us.
Q: Can we put our logo on your products?
A: Yes, we can customized your logo. But the quantity of the items have special requirements.
Q: What certifications do products have for your company?
A: CE, ISO and SGS at present.
Q: Do you provide technical support?
A: Yes, technical support is available after sales,if needed,we will provide our technicist to overseas to service you.